The range is designed around a scalable architecture and offers the following features as standard:
- Optimum chamber environment, profiled heating creates gentle circulation – no fan or HEPA filter.
- Seamless chamber, with rounded interior corners makes cleaning easy.
- Removable shelves are non-tip.
- Sealed inner glass door allows chamber viewing.
- CO2 range programmable from 0.2 to 20 %.
- Temperature range programmable from 4°C above ambient to 50°C.
- Infrared CO2 Sensor, with auto self-referencing function.
- Large, easy to read LCD display with LED backlight.
- Removable, seamless, stainless steel humidity reservoir.
- 25mm access port for cable pass-through etc.
- HEPA filtration for all incoming gasses.
- Large range of options allows you to configure each incubator for each task.
- Optional data capture and reporting for QA and verification purposes.
- 3 Year Warranty (extendable).
Control System Safety Features Include:
- Independent over temperature cut-out.
- Calibration and programmed data internally backed up to guard against loss.
Oxygen Control
An Oxygen control options are available: 1-19%
The control systems consist of nitrogen injection to accurately control the desired below ambient Oxygen level.